$ 1578.99
SKU:
GE883470009961
Category:
Infrared light emitted from the source interacts with suspended particles in the sample, and the resulting scattered light is detected by a photodetector. This is then converted to an electrical signal, processed to yield an accurate turbidity reading through analog and digital signal processing.
Applications:
Key Features:
Ordering Notes:
1. All Sper Inline Water Quality Meters and Sensors are sold separately – one of each is required for a complete system.
2. All Inline Water Quality Meters and Sensors ship direct from the factory. Please allow 10-15 business days for shipping and handling.
| Measuring Range | 0.01-100 NTU0.014000 NTU |
| Accuracy | Less than the measured value of 1%, or 0.1NTUwhichever is larger |
| Pressure Range | 0.4Mpa |
| Current Speed | 2.5m/s, 8.2ft/s |
| Calibration | Sample calibration, slope calibration |
| Sensor Main Material | Body: SUS316L + PVCnormal type), SUS316L Titanium + PVCsea water type)O type circleFluorine rubber cablePVC |
| Power Supply | 12V |
| Communication Interface | MODBUS RS485 |
| Temperature Storage | 5 to 149 F (-15 to 65) |
| Working Temperature | 32 to 113F (0 to 45) |
| Size | 2.3 x 10 inches (60mm x 256mm) |
| Weight | 3.6lbs. (1.65kg) |
| Protection Grade | IP68/NEMA6P |
| Cable Length | Standard 10m cable, can be extended to 100m |
Instruction Manual